2 recalls Categories: Surgical Instruments

Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.

Nov 11, 2014 Surgical Instruments Nationwide View Details →

The firm initiated the product recall because labeling on some of the intermediate boxes of catheters identify the contents to be M16C, a coud¿-tip catheters while the actual contents are M16Fr male, 16Fr, straight-tip catheters with the pouch and the case box labeled as M16Fr male, 16Fr, straight-tip catheters.

Sep 19, 2012 Surgical Instruments Nationwide View Details →