Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.
The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.
Updated Manual-Warnigs have been added stating that exceeding 40'C for extended periods may cause tissue damage
Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration on the board due to overheating and shut down of the device.
Potential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended change to control settings.