Biomeme

3 recalls • • Categories: Diagnostic Equipment

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).

Oct 20, 2021 Diagnostic Equipment View Details →

Users cannot complete testing due to a sign-error in the software component that controls the filter movement and results in an instrument failure and assay failure before results are generated.

Jan 5, 2021 Diagnostic Equipment View Details →

The firm has become aware of nine reports by a single customer that the software made false positive calls. The investigation determined that the root cause is the higher than expected amber-to-red co-excitation in the assay.

Oct 23, 2020 Diagnostic Equipment Nationwide View Details →