American Optisurgical

4 recalls • • Categories: Surgical Instruments

American Optisurgical, Inc. is recalling certain lots of TX1 Tissue Removal System disposable handpiece due to inventory being released into distribution without a cleared FDA 510(k) to cover recent product changes. Use of this product may lead to an underperformance of the device.

Apr 30, 2013 Surgical Instruments Nationwide View Details →

The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, Inc. will be making a correction to add spacers between the valve and cassette to correct a small irrigation leak that can occur after the priming cycle. Use of this product may cause the device to lose power and stop wo

Mar 30, 2013 Surgical Instruments View Details →

Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is incorrect.

Feb 22, 2012 Surgical Instruments Nationwide View Details →