Alere San Diego

49 recalls Categories: Other Medical Devices, Diagnostic Equipment, Implants & Prosthetics, Infusion Pumps

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

Jun 24, 2022 Diagnostic Equipment Nationwide View Details →

The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.

Feb 15, 2019 Diagnostic Equipment Nationwide View Details →

The recalled lots have demonstrated unexpected false positive THC results.

Nov 14, 2017 Diagnostic Equipment Nationwide View Details →

Encoded with incorrect ranges that are not consistent with the ranges printed on the Expected Value Card.

Oct 12, 2016 Diagnostic Equipment Nationwide View Details →

Incorrect expiration date on the magnetic strip that does not match the labeled expiration date.

Aug 2, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Jul 11, 2016 Diagnostic Equipment Nationwide View Details →

Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect.

Aug 27, 2015 Diagnostic Equipment Nationwide View Details →

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

Dec 5, 2014 Diagnostic Equipment Nationwide View Details →

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

Dec 5, 2014 Diagnostic Equipment Nationwide View Details →

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

Dec 5, 2014 Diagnostic Equipment Nationwide View Details →

Alere San Diego is recalling Alere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems because the bottle labeled as Calibrator Level 5 actually contains Calibrator Level 4. This results in a calibration failure when attempting to calibrate using a calibrator kit which contains the affected Level 5 Calibrator.

May 30, 2014 Diagnostic Equipment Nationwide View Details →

Alere initiated this recall because a limited number of Alere Triage¿ D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.

May 2, 2014 Diagnostic Equipment Nationwide View Details →

These lots may not recover within range when tested using certain commercially available controls.

Mar 20, 2014 Diagnostic Equipment Nationwide View Details →

Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Apr 16, 2014 Diagnostic Equipment Nationwide View Details →

Alere San Diego, Inc. is initiating a recall of the Alere Triage¿ TOX Drug Screen, Catalog Number 94400, Lots W56013B and W56035B because Alere's investigation has shown that these lots have a frequency of false negative results for PCP (Phencyclidine) that is higher than the frequency shown in the package insert. If you have questions about previously reported results.

Feb 5, 2014 Diagnostic Equipment Nationwide View Details →

Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.

Jun 3, 2013 Diagnostic Equipment View Details →

Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot # number printed was 321067.

Oct 23, 2013 Diagnostic Equipment Nationwide View Details →

Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.

Sep 23, 2013 Implants & Prosthetics Nationwide View Details →

The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for

Apr 16, 2013 Diagnostic Equipment Nationwide View Details →

Alere San Diego Inc. is recalling the Cardinal Health hCG Cassette Rapid Test because some of them were incorrectly packaged into Cardinal Health hCG Combo Rapid test kit boxes. The kit box incorrectly indicates the product may be used for both urine and serum sample types.

Jan 21, 2013 Diagnostic Equipment Nationwide View Details →

The Alere Cholestech LDX¿ hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the same patient sample.

Mar 6, 2013 Diagnostic Equipment Nationwide View Details →

Alere San Diego is recalling the Clearview Mononucleosis Cassette because the positive control included in the rapid test may produce false negative results.

Feb 26, 2013 Diagnostic Equipment Nationwide View Details →

Out of range (increased) HDL Cholesterol results.

Oct 22, 2012 Diagnostic Equipment Nationwide View Details →

Out of range (increased) HDL Cholesterol results.

Oct 22, 2012 Diagnostic Equipment Nationwide View Details →

Biosite is recalling the Triage TOX Drug Screen kits where the box containing test devices because it was mislabeled.

May 28, 2009 Diagnostic Equipment Nationwide View Details →

Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.

Jul 12, 2012 Diagnostic Equipment Nationwide View Details →

A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.

May 14, 2012 Diagnostic Equipment Nationwide View Details →

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Infusion Pumps Nationwide View Details →

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Other Medical Devices Nationwide View Details →

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Other Medical Devices Nationwide View Details →

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Infusion Pumps Nationwide View Details →

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

May 22, 2012 Infusion Pumps Nationwide View Details →