A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.
AGFA Healthcare
A customer experienced when using IMPAX CV Reporting software, specifically, when building a NIV report, the NIV Cardio report was showing incorrect findings for Aneurysm.
One of the dent markers from the DX-D600 dropped down from the overhead rail.
The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.
Customers have experienced IMPAX CV software, specifically, Demographics Manager (DM), not producing correct pdf reports after the reports, at the time of physician signature, were correct.
Firm received a complaint that the hospital had sent two accession numbers over the same morning and on the Agility side the patient name was incorrect.
When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur.
Some material of the coating Type CXCPMV3 has potential artifacts showing a small bright thin line when the film is processed.
Unit was mislabeled with a factory label showing 32 KW output power.
Some sheets of the medical screen film are fogged.
Some customers reported that while using the DX-D600 in room or exam preparation, the Overhead Tube Crane (OTC) showed increased velocity and uncontrolled longitudinal movement.