Distribution without Pre-Market Clearance
Acumed
Due to potential breakage during use.
Distribution without Pre-Market Clearance
Distribution without Pre-Market Clearance
Distribution without Pre-Market Clearance
Distribution without Pre-Market Clearance
Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture, joint impingement or damage, or soft tissue damage.
Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture, joint impingement or damage, or soft tissue damage.
The firm initiates a voluntary removal and discontinue marketing of Slide-Loc Anatomic Radial Head System based on its performance in the market. The firm received product complaints regarding instrumentation performance, trial product design, and implant dissociations.
Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).
The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.
OPT Croutons, part number 65-0210-S, is recalled due to the result of inadequate documentation of the sterilization dose used in production. Product sterilized at the site did not receive the minimum dose required per the new performance qualification.
Manufacturing of these devices with a grade of stainless steel that is not within specifications.
Certain lots of Acumed Tension Band Pins are being recalled because the implantable product was manufactured with a grade of stainless steel that is not within specifications.