Acist Medical Systems

3 recalls Categories: Diagnostic Equipment, Surgical Instruments, Infusion Pumps

Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence.

Jan 22, 2021 Diagnostic Equipment Nationwide View Details →

Reports related to ingress of air into the manifold.

Dec 21, 2015 Surgical Instruments View Details →

ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w

Mar 15, 2013 Infusion Pumps Nationwide View Details →